Helium Leak Detectors and Leak Testing Systems

A helium leak detector uses a mass spectrometer tuned to helium to find and quantify gas losses from a vacuum system or a pressure vessel. It is the reference method for very tight seals, from checking a vacuum chamber to series testing of components in production, where the bubble test or the pressure-decay test is no longer sensitive enough.

The principle is simple: helium, a light, inert gas present in air in very small quantity, is sprayed on the outside of a system under vacuum or introduced inside a vessel, and the spectrometer detects the amount passing through the leak. The leak rate is expressed in mbar·l/s or Pa·m³/s and shows how much gas passes per unit of time. In the vacuum method, the test object is connected to the detector, which evacuates it and measures the helium that gets in; in the sniffer probe method, the object is filled with helium under pressure and a hand-held probe searches for the leak on the outside.

A test system can be a portable detector for service, a bench-top model with integrated pumping, or an automated station with a test chamber, used in component production. The selection criteria include minimum sensitivity, response and clean-down time, mobility, calibration with a reference leak and the interface with the production line. Tracer gas test methods are described in standards, for example ISO 20485.

How to choose

Sensitivity required

Determine the maximum leak rate permitted for the product or system; the detector must have a minimum detectable threshold better than this value, with a margin for the helium background.

Test method

The vacuum method is more sensitive, while the sniffer probe method is suitable for pressure vessels or systems that cannot be put under vacuum.

Response and clean-down time

A detector that cleans down quickly after a large signal reduces the time lost in locating multiple leaks or in series testing.

Pumping capacity

For large volumes, the integrated pump must reach the test pressure quickly; otherwise an external auxiliary pump is used.

Portable versus fixed installation

For field service a portable, rugged and easily transported unit is useful; for production, an integrated station with automation and a test chamber.

Calibration and reference leak

The detector is calibrated with a standard leak; check whether the instrument includes an internal one and how easily calibration can be redone in service.

Interface and reporting

For production lines, check communication with the control system, the pass/reject criteria and result storage for traceability.

What to send for a quotation

  • The test object, its volume and working pressure
  • The maximum permitted leak rate (in mbar·l/s or Pa·m³/s)
  • The preferred method: vacuum, sniffer probe or both
  • The environment of use: laboratory, workshop, field or production line
  • The test cadence, if this is series production
  • The model of the existing detector, if you are asking for consumables or service

Typical applications

  • Checking the tightness of vacuum chambers and lines
  • Testing components in production
  • Inspection of refrigeration and air-conditioning systems
  • Checking pressure vessels and pipework

Maintenance and spare parts

The spectrometer and the pumps of the detector need checks according to the manufacturer's schedule, and the ion source filament is a wear part that is replaced periodically. Calibrate the instrument with the standard leak before critical measurements and periodically check the validity of the reference leak. Avoid helium contamination of the workplace, because a high background limits the effective sensitivity of the test, and keep the usual spare parts at hand so that the instrument does not sit idle during production periods.

Brands we quote for this product type

Frequently asked questions

Why is helium used for leak detection?

Helium has very small molecules, so it passes through minute leaks, it is inert and does not react with the materials tested, and air contains only a very small amount of it, which gives a low background. It is easily detected by mass spectrometry, with fast response and high sensitivity. That is why the method is used for tightness levels at which bubble tests, pressure-decay tests or tests with other gases do not have the sensitivity required.

What is the difference between vacuum testing and sniffer probe testing?

In vacuum testing, the object is connected to the detector and evacuated, and helium sprayed on the outside or present in a test chamber enters through the leak and is detected; the method is very sensitive and allows quantitative measurement. In probe testing, the object is filled with helium and an operator moves the probe over the surface to locate the leak. The second method is less sensitive, but it also applies where the object cannot be put under vacuum.

What information do I need for a quotation for a leak detector?

Tell us what you are testing, the volume and working pressure, the maximum acceptable leak rate, the preferred method and the environment in which the instrument will work. For series testing, add the cadence and the requirements for integration with the line. Lead time is usually 24–72 h when the item is in stock, locally or at the supplier, and 1–4 weeks when ordered in; depending on the brand and specification it can take longer.

How often should a helium leak detector be calibrated?

Calibration with the standard leak is normally carried out before every series of important measurements or at the interval set by the quality procedure, and the reference leak itself is checked periodically against a standard. The frequency depends on the requirements of the application and the manufacturer's recommendation. In series production, a daily or per-shift check with a standard is the usual practice, to support the validity of the results.