Skip to content
+40 371 232 404|Mon - Fri: 08:00 - 16:30 (Romania time)

Turbomolecular and High Vacuum Pumps

A turbomolecular pump reaches pressures in the high and ultra-high vacuum range by transferring momentum from a bladed rotor, spinning at very high speed, to the gas molecules. It does not exhaust directly to atmosphere, so it always works together with a backing pump, and it is used on process chambers, mass spectrometers, electron microscopes and thin-film deposition systems.

A turbomolecular pump consists of several stages of rotors and stators with inclined blades. At very low pressures, when gas molecules no longer collide with each other but only with the walls, the rotor blades give them a directed motion towards the exhaust. The compression ratio depends strongly on molecular mass: it is very high for heavy gases and much lower for hydrogen and helium, which is why the predominant gas in the process influences the ultimate pressure. The pump cannot start from atmospheric pressure and needs a backing pump to keep the backing pressure within the permitted limits.

When selecting, what matters is the pumping speed, expressed in litres per second for a given gas, the ultimate pressure, the type of bearings and the inlet flange. Hybrid bearings, with ceramic balls and lubrication, are the most widespread, while magnetically levitated bearings eliminate the lubricant and reduce vibration, an advantage in microscopy and in contamination-sensitive processes. The controller, air or water cooling and protection against sudden air inrush complete the selection.

How to choose

Pumping speed for the process gas

Check the effective speed for your own gas, not only for nitrogen; for hydrogen and helium the value is usually significantly different, and the chamber must be sized for the real gas load.

Ultimate pressure required

Determine whether you work in high or ultra-high vacuum; for the latter, metal flanges, bake-out and a suitable backing pump are needed, not just a high-performance turbomolecular pump.

Bearing type

Hybrid bearings are economical and robust; magnetic levitation offers lubricant-free operation and low vibration, useful in electron microscopy and analytical applications.

Inlet flange and mounting orientation

The inlet must match the chamber standard (ISO-K, ISO-F or CF), and the mounting position permitted by the manufacturer is checked before the installation is designed.

Compatibility with corrosive gases or particles

Processes with aggressive gases or deposits call for versions with special protection, body heating and, sometimes, a trap upstream of the pump.

Associated backing pump

The capacity of the backing pump must keep the backing pressure below the limit of the turbomolecular pump at the maximum gas flow, not only during the initial pump-down of the chamber.

Controller, cooling and interface

Determine whether a separate or integrated controller is needed, air or water cooling, and communication with the plant control system.

What to send for a quotation

  • The gas or gas mixture in the process and the estimated flow
  • The required ultimate pressure and the working pressure
  • The chamber inlet flange (type and nominal size)
  • The mounting position and the available space
  • The model of the existing backing pump, if there is one
  • The make and code of the pump being replaced, if this is a replacement

Typical applications

  • Mass spectrometry and surface analysis
  • Thin-film deposition and sputtering
  • Electron microscopy
  • Research facilities and space simulation

Maintenance and spare parts

The service interval depends on the type of bearings and the degree of process contamination. With hybrid bearings, the bearings and the lubricant reservoir are replaced at the intervals indicated by the manufacturer or after a given number of operating hours. Avoid sudden venting of the chamber while the rotor is still turning fast, and monitor the temperature and the motor current, because a slow rise signals wear or deposits.

Brands we quote for this product type

Frequently asked questions

Why does a turbomolecular pump need a backing pump?

The rotor of a turbomolecular pump transfers molecules only when their passage is free, that is, at very low pressures; at pressures close to atmospheric the blades would overheat and be mechanically overloaded. The backing pump performs the initial pump-down of the chamber to a level at which the turbopump can start, and then takes the gas exhausted by it, keeping the backing pressure below the permitted limit. Without it, neither start-up nor continuous operation is possible safely.

Are magnetic bearings worth it compared with hybrid ones?

It depends on the application. Magnetic bearings have no mechanical contact, so they need no lubricant, produce very little vibration and introduce no oil vapour into the chamber, which matters in electron microscopy and in contamination-sensitive processes. On the other hand, the solution is electronically more complex. For most general-purpose installations, hybrid bearings offer a good balance of cost, service life and simple maintenance.

What information should I send for a quotation for a turbomolecular pump?

State the process gas and the estimated flow, the ultimate and working pressure, the chamber inlet flange, the mounting position and the backing pump used. If you are replacing an existing model, send the make and code from the pump nameplate. Lead time is usually 24–72 h when the item is in stock, locally or at the supplier, and 1–4 weeks when ordered in; depending on the brand and specification it can take longer.

Can I mount the turbomolecular pump in any position?

Not always. Some models allow any orientation, others have restrictions related to the bearings, the circulation of the lubricant or the drainage of condensate, and for a vertical position with the inlet facing down or for side mounting, the mounting position permitted for the chosen model must be checked. In addition, the fixing structure must withstand the braking torque if the rotor seizes, a rare case but one that requires correct anchoring of the flange.

Updated: 2026-11-20